Abstract:
Objective LA−ICP−MS in situ microanalysis is significantly affected by matrix effects, thus requiring matrix−matched external standards for calibration. This study aims to develop a new zircon U−Pb dating reference material—zircon HF−1.
Methods Zircon HF−1 was collected from trachyandesite in Gaize County, Tibet, and subjected to LA−ICP−MS microanalysis. The 91500 standard was used to calibrate unknown sample ages, while zircon Plešovice was simultaneously analyzed as a monitor. The performance of zircon HF−1 was compared with that of the 91500, Plešovice, Qinghu, and Penglai standards.
Results A total of 529 spot analyses were obtained in five groups. The weighted mean 206Pb/238U age for all spots is 238.35 ± 0.31 Ma (MSWD = 0.117, n = 529), with a corresponding 206Pb/238U ratio of 0.037675 ± 0.000052 (MSWD = 0.104). The weighted mean 206Pb/238U age of the Plešovice standard is 337.4 ± 1.3 Ma (MSWD = 0.086, n = 36), consistent with the reported value of 337.4 ± 1.0 Ma (2σ, n = 68), confirming the accuracy and reliability of the analytical data.
Conclusion Zircon HF−1 exhibits uniform U−Pb age, abundant availability, and ease of acquisition, with an age of 238 Ma. It can fill the gap in zircon reference materials for the early Mesozoic (~240 Ma) and has potential as a substitute for zircon Plešovice in internal laboratory monitoring. However, further research is needed to evaluate its Hf isotopic homogeneity.